Advanced settings


The SensAlert 300 is supplied with a standard setting of a 2 second delay
(the person must leave the bed/chair for 2 seconds before alarm triggers)
with TONE Alert set to MEDIUM Volume.


• Select Alert option (Factory default is TONE Only)
o Tone
o Voice
o Voice & Tone

• Recording A message for VOICE Alert
o Press & Hold REC Button
o Speak into MIC
o Release button when message is complete

• Select Alert Volume (Factory default is MEDIUM Volume)
o Low
o Medium
o High

• Mute Function for use with Nursecall Systems (Factory default is MUTE OFF)
o ON (Silent MODE)
o OFF (Tone/Voice Alert active)

• Time Delay Options
Used to provide an “out of bed too long” alert, where an individual is capable of going to the bathroom themselves at night, but may be at risk of occasionally falling.
This function will allow the user to exit the bed without triggering an alarm unless the user does not return to bed within the delay time.



• Power
o Battery (3x AAA) included
o Optional Mains Adapter
o The LED on the front of the control unit also functions as a low battery indicator.
When the batteries are low and require changing the light bar will flash blue intermittently
and the alarm will emit a beeping sound.